Measurement-computing CB-COM-7012 Manual do Utilizador Página 64

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 72
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 63
64
CB-7012, CB-7014, CB-7017 Users Manual
2.41 ~AA4
Description: Read PowerOn Value and Safe Value
Note: This command is for CB-7012/12D/12F/12FD/14D only.
Syntax: ~AA4[CHK](cr)
~ Delimiter character
AA Address of module to read (00 to FF)
4 Command for reading PowerOn Value and Safe Value
Response: Valid Command: !AAPPSS[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax or communication error may get no response.
! Delimiter for valid command
? Delimiter for invalid command
AA Address of responding module (00 to FF)
PP PowerOn Value, 00=DO0 off, DO1 off, 01=DO0 on,
DO1 off, 02=DO0 off, DO1 on, 03=DO0 on, DO1 on
SS Safe Value, 00=DO0 off, DO1 off, 01=DO0 on, DO1
off, 02=DO0 off, DO1 on, 03=DO0 on, DO1 on
Example:
Command: ~014 Receive: !010000
Read Address 01 PowerOn/Safe Value, return PowerOn Value
is DO0 off, DO1 off, Safe Value is DO0 off, DO1 off.
Related Command:
Sec. 2.42, ~AA5PPSS
Related Topic:
Set. 3.2, Module Status; Sec. 3.3, Dual Watchdog Operation