2.5 $AA0
Description: Perform Span Calibration
Syntax: $AA0[CHK](cr)
$ Delimiter character
AA Address of module to set (00 to FF)
0 Command for performing span calibration
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax or communication error may get no response.
! Delimiter for valid command
? Delimiter for invalid command
AA Address of responding module (00 to FF)
Example:
Command: $010 Receive: !01
Perform Address 01 span calibration, return successful.
Command: $020 Receive: ?02
Perform Address 02 span calibration; return calibration not
enabled before performing calibration command.