USB-TEMP-AI User's Guide Specifications
26
Note 19: All ground pins are common and are isolated from earth ground. If a connection is made to earth
ground when using both digital I/O and conductive thermocouples, the thermocouples are no longer
isolated. In this case, thermocouples must not be connected to any conductive surfaces that may be
referenced to earth ground.
Counter
Table 16. CTR I/O specifications
TTL, rising edge triggered
Counter read/writes rates
(software paced)
Counter read: System dependent, 33 to 1000 reads per second.
Counter write: System dependent, 33 to 1000 reads per second.
Schmidt trigger hysteresis
4.0 V min, 5.5 V absolute max
1.0 V max, –0.5 V absolute min
Note 20: All ground pins are common and are isolated from earth ground. If a connection is made to earth
ground with both the counter (CTR) and conductive thermocouples, the thermocouples are no longer
isolated. In this case, thermocouples must not be connected to any conductive surfaces that may be
referenced to earth ground.
Memory
Table 17. Memory specifications
1,024 bytes isolated micro reserved for sensor configuration
256 bytes USB micro for external application use
Microcontroller
Table 18. Microcontroller specifications
Two high-performance 8-bit RISC microcontrollers
USB +5V voltage
Table 19. USB +5V voltage specifications