USB-TEMP-AI User's Guide Specifications
25
Settling time is defined as the time required for a channel to settle within a specified accuracy in response to a
full-scale (FS) step input.
Analog input calibration
Table 13. Analog input calibration specifications
+10.000 V, ±5 mV max. Actual measured values stored in EEPROM
Long term stability: 30 ppm/1000 h
Throughput rate
Table 14. Throughput rate specifications
2 S/s on each channel, 4 S/s total
2 S/s on each channel, 6 S/s total
2 S/s on each channel, 8 S/s total
2 S/s on each channel, 10 S/s total
2 S/s on each channel, 12 S/s total
2 S/s on each channel, 14 S/s total
2 S/s on each channel, 16 S/s total
Note 18: The analog inputs are configured to run continuously. Each channel is sampled twice per second.
The maximum latency between when a sample is acquired and the voltage/temperature data is
provided by the USB unit is approximately 0.4 seconds.
Digital input/output
Table 15. Digital input/output specifications
Independently configured for input or output.
Power on reset is input mode unless bit is configured for alarm.
Pull-up/pull-down
configuration
All pins pulled up to +5 V via 47 K resistors (default). Contact MCC factory for
pull-down to ground (GND) capability.
Digital I/O transfer rate
(software paced)
Digital input – 50 port reads or single bit reads per second typ
Digital output – 100 port writes or single bit writes per second typ
2.0 V min, 5.5 V absolute max
0.8 V max, –0.5 V absolute min
Output low voltage
(IOL = 2.5 mA max)
Output high voltage
(IOH = –2.5 mA max)