Chapter 3
Storing Device Data on the PC
When you select View Devices»On This PC and connect a device to a WiFi network, the GENERAL
SETTINGS dialog box displays with general device and logging settings for the a WiFi-500 Sensor Series data
logger
Changing device logging settings
Complete the following steps to change device logging settings from the PC:
1. Enter the following information In the
GENERAL SETTINGS dialog box:
o
Name Device to give it a unique identity: Enter a unique name to identify the device on the WiFi
network.
o Select temperature scale: Click °C or °F.
o Select sample rate: Select a sample rate between 10 Seconds and 12 Hours from the drop-down list.
This value sets how often the device takes a reading.
o
Select how often the device will communicate with the PC: Select how frequently between every
1 Minute to every 24 hoursthe device transmits samples to the host computer
For example, if the sample rate is set to
10 Seconds and the sample transmission frequency is set to
1 Minute, the device transmits six samples to the computer every minute (60 seconds ÷ 10 seconds = 6
samples).
Figure 3: GENERAL SETTINGS dialog box
The more frequently data is transmitted the quicker battery power is depleted. By increasing the time
between transmissions, the battery charge lasts longer.
2. Click
Next.