Measurement-computing PCM-DAS16x/16 Manual do Utilizador Página 30

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 40
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 29
6.2.7 Isolated Grounds / Differential Inputs
Optimum performance with isolated signal sources is assured with the use of the dif-
ferential input setting. The diagram below shows the recommend connections is this
configuration..
26
+
-
Input
Amp
To A /D
A/D Board
I/O
Connector
LL GND
CH High
CH Low
Signal Source
and A/D Board
Already Isolated.
Already isolated signal source
and A/D board connected to
a differential input.
GND
These grounds are
electrically isolated.
10 K
10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your system’s noise immunity.
Vista de página 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 39 40

Comentários a estes Manuais

Sem comentários