Service Request Mask (SRQ) Mn
The
Service Request
(SRQ)
mode is used by the
Digital488
to alert the
controller to one of several conditions described below. Multiple
SRQ
conditions can
be enabled simultaneously by issuing them sepatately or by combining them in one
command. If multiple
SRQ
commands are contained within the same command string,
each
SRQ
command should be followed by an
Execute
command
(X).
The
programmed
SRQ
modes will remain enabled until the
M0
command is sent, or the
controller sends a Device Clear (DCL), Selected Device Clear (SDC), or Interface Clear
(IFC) command.
M0
SRQ is disabled
M1
SRQ on
Service
input transition
M2
SRQ on
EDR
input transition
M4
SRQ on bus error
M8
SRQ on Self-Test error
M16
SRQ on Ready
MO
default mode disables the
SRQ
function, preventing the
Digital488
from
generating a
Service Request
.
M1
will generate a
Service Request
when the
Service
Input line makes a
transition. Refer to the
Invert
command
(I64)
description for programming the polarity
of the
Service
input line.
M2
will generate a
Service Request
when the
EDR
input
makes a transition.
Refer to the
Invert
command
(I32)
description for programming the polarity of the
EDR
input line.
M4
will generate a
Service Request
when a bus error occurs. The most
common bus error is sending an invalid command to the
Digital488
. For example,
attempting to select an 'F6' format when no 'F6' format exists will generate a
Service
Request
when the
M4
mode is selected.