Section 2 Getting Started
2.14
into the unit. If the problem is unresolved, refer to the
Service Information
section
of this manual.
Connect a serial cable to the DB-25 connector on the rear of the
Digital232
.
Connect the other end of the cable to the serial host. Running a dumb terminal
program similar to that shown in Appendix C, type the following....
@
Reset the
Digital232
R0
Request data from the
Digital232
The
Digital232
should respond with '
FFFFFFFFFF'
2.5 Errors
The
Digital232
has built-in error checking, to ensure that it has received valid
commands and data, and to alert the user if an inappropriate situation has occurred.
The front panel
Error
LED will illuminate when an error condition occurs, and will
remain lit until the status command
U0
is received. The
Digital232
will continue to
operate after an error has occurred, but in the instance of an invalid command, the
command will not be executed and must be re-sent.
The
Service Request
command
M4
can be used to automatically send an SRQ
message to the host whenever an error occurs. This is particularly helpful when first
configuring the system using a 'dumb terminal' program, as described for the IBM PC
in Appendix C. Refer to the
Status
command (
U0
) for more information on error
codes.