Measurement-computing CIO-EXP32 Manual do Utilizador Página 21

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 27
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 20
CIO-EXP32 User's Guide Functional Details
Output voltage vs gain at maximum temperature
Type
Min Temp °C Max Temp °C
mV at
Max
Gain / Output Volts @ Max Temp
100 200 300 500 800
J 0 750 42.28 4.2 8.5 12.7 21.1 33.8
K
200
1250 50.63 5.1 10.1 15.2 25.3 40.5
T
200
350 17.82 1.8 3.6 5.3 8.9 14.3
E
200
900 68.78 6.9 13.8 20.6 34.4 55.0
S 0 1450 14.97 1.5 3.0 4.5 7.5 12.0
R 0 1450 16.74 1.7 3.3 5.0 8.4 13.4
The table below lists the maximum readable temperatures for each gain and thermocouple type for an A/D
range of ±5V.
Gain vs maximum readable temperatures
Type Gain / Max. Readable Temps. with A/D Range = ±5 V
10 100 200 300 500 800
J MAX MAX 457 °C 307 °C 186 °C 117 °C
K MAX 1232 °C 602 °C 406 °C 246 °C 152 °C
T MAX MAX MAX 330 °C 214 °C 140 °C
E MAX 661°C 350 °C 244 °C 152 °C 98 °C
S MAX MAX MAX MAX 1035 °C 697 °C
R MAX MAX MAX 1445 °C 961°C 658 °C
To read the maximum usable temperature for type K and E thermocouples, set the gain to X10 or less.
After you change the gain setting, calibrate the CIO-EXP32 using the InstaCal utility.