USB-2404-10 User's Guide Specifications
17
Safety voltages
Table 7. Safety specifications (Note 3)
Channel-to-earth ground isolation
250 Vrms, Measurement Category II (Note 4)
2,300 Vrms, verified by a 5 sec dielectric withstand test
Channel-to-channel isolation
250 Vrms, Measurement Category II (Note 4)
1390 Vrms, verified by a 5 sec dielectric withstand test
Note 3: Connect only voltages that are within the limits specified in this table.
Note 4: Measurement Category II is for measurements performed on circuits directly connected to the
electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example 115 V for US or 230 V for Europe.
Caution! Do not connect the device to signals or use for measurements within Measurement Categories
III or IV.
Screw terminal connectors
Table 8. Screw terminal connector specifications
16 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped
from the end.
Torque for screw terminals
0.22 to 0.25 N· m (1.95 to 2.21 lb. · in.)
Table 9. Screw terminal assignments
Accessory products
Table 10. Screw terminal connector specifications
Two-position detachable screw terminal connector blocks (quantity ten)
Backshell for use with the ACC-102 two-position screw terminal connector blocks. Provides
strain relief and operator protection from high-voltage signals (quantity four)