Measurement-computing DBK70 Manual do Utilizador Página 29

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 28
DBK70 User’s Manual 919092 Hardware Reference 4-11
Analog Output Cards
The Analog Output Cards (DBK70-P4) are inserted into the 4 sockets adjacent to the analog output
connectors on the DBK70. Populate the sockets starting with the right-most socket continuing left until the
desired number of cards have been installed. Make sure no card has any open sockets to its right. The cards
must be oriented as shown in the diagram below.
When the DBK70 powers-up, the DBK70-P4s are immediately recognized and initialized. Each DBK70-P4
carries its own calibration constants on an on-board memory device so cards can be inserted in the field
without the necessity of re-calibration.
Add more cards
in this direction
The 1
st
card
goes here
Insert as
shown
Vista de página 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 107 108

Comentários a estes Manuais

Sem comentários